BD Bactec FX40

BD BACTEC FX40 Bacteriological Analyzer

Registration Certificate in RK: RK-MT-7 No. 014761

Registration Date: September 11, 2015


System Overview

The BD BACTEC FX40 is an automated system designed for the rapid detection of bacteria and fungi in clinical specimens.+1

Key Features:

  • Modular Design: Offers installation flexibility to meet specific laboratory requirements.
  • User Interface: Features a color touch-screen graphic display for ease of use.
  • Real-time Monitoring: Internal sensors immediately detect the placement or removal of vials.+1
  • Versatility: Supports simultaneous processing of bacterial, fungal, and mycobacterial cultures using various media types.
  • Connectivity: Can be integrated with a compatible Laboratory Information System (LIS) or the BD EpiCenter data management system.

Method Principles

  1. Metabolism: Microorganisms in the culture vials consume nutrients and release carbon dioxide ($CO_2$) into the medium.
  2. Reaction: A dye in the sensor at the bottom of the vial reacts with the $CO_2$.
  3. Detection: This reaction modulates the light absorbed by the detector’s fluorescent material.
  4. Measurement: A photodetector at each station measures fluorescence levels every ten minutes.
  5. Alerts: When a “positive” result is identified based on pre-programmed parameters, the system triggers an indicator light, an audible alarm, and displays the culture on the tablet.
  6. Confirmation: Laboratory technologists then remove positive vials for subculturing and identification.

Media Overview

The system supports several specialized media types:

Medium TypeCatalog No.Recommended Volume (Optimal)Key Features
Standard/10 Aerobic /F4422603.0 – 10.0 ml (8.0 – 10.0 ml)General aerobic testing.
Plus Aerobic /F4421923.0 – 10.0 ml (8.0 – 10.0 ml)Contains resins for antibiotic neutralization.
Standard Anaerobic /F4421913.0 – 7.0 ml (5.0 – 7.0 ml)General anaerobic testing.
Peds Plus /F4421940.5 – 5.0 ml (1.0 – 3.0 ml)Optimized for pediatric/low-volume samples; includes resins.
Plus Anaerobic /F4421933.0 – 10.0 ml (8.0 – 10.0 ml)Anaerobic medium with antibiotic-neutralizing resins.
Lytic/10 Anaerobic /F4422653.0 – 10.0 ml (8.0 – 10.0 ml)Contains saponin to lyse blood cells and release phagocytized organisms.
Myco/F Lytic4420031.0 – 5.0 ml (3.0 – 5.0 ml)Specialized for yeast, fungi, and mycobacteria.+1
Mycosis IC /F442026/063.0 – 10.0 mlSelective medium for fungal recovery.

Technical Specifications

SpecificationIndividual InstrumentWith Rack
Height39.1 cm77.5 cm
Width67.5 cm67.5 cm
Depth58.5 cm58.5 cm
Weight (Empty/Full)31.8 kg / 38.0 kg64.6 kg / 76.9 kg

Operating Requirements:

  • Voltage: 100 – 240 VAC ± 10% (50 – 60 Hz).
  • Operating Temperature: 18.0°C to 30.0°C.
  • Humidity: 25% to 80% non-condensing.
  • Placement: Level surface, away from direct sunlight, heat sources, or excessive dust.
  • Noise Level: ≤ 55 dBA at 1 meter.

Note: This information is an overview. A full electronic copy of the User Manual is available upon request.